Department of Botany, Marsh Life Sciences Building,
University of Vermont, Burlington, Vermont 05402 (C.W.); United States
Department of Agriculture Forest Service, 705 Spear Street, Burlington,
Vermont 05402 (M.T.T.); and Lehrstuhl für Pflanzenökologie,
Universität Bayreuth, UniversitätStrasse 30, D-95440
Bayreuth, Germany (E.S.)
 |
INTRODUCTION |
In recent years, the cohesion-tension (CT) theory of the ascent of
sap in plants has been questioned (Balling and Zimmermann, 1990
;
Benkert et el., 1991
; Zimmermann et al., 1993
). According to direct
measurements with cell pressure probes, the pressures measured in xylem
vessels (Px) were usually not more
negative than
0.5 MPa (reference zero pressure = atmospheric
pressure). Indirect measurements of Px
using a Scholander-Hammel pressure bomb (Scholander et al., 1965
)
suggest that Px might be as low as
10 MPa
in some plants (Kolb and Davis, 1994
; Kramer and Boyer, 1995
; Steudle,
1995
; Sperry et al., 1996
; Tyree, 1997
). Early pressure probe results
have failed to corroborate the pressure bomb (Balling and Zimmermann,
1990
) although better agreement has been reported more recently
(Melcher et al., 1998
). Nevertheless, prior reports have failed to
emphasize that the validity of the CT theory is independent of the
correctness of pressure-bomb estimates of
Px (Tyree, 1997
). Failure of the
pressure probe to detect rapid changes in
Px following rapid changes in
transpiration (Benkert et al., 1991
) has provided a more troubling
conflict with the CT theory.
The CT theory, as originally proposed (Dixon and Joly, 1894
), makes few
quantitative predictions of how negative
Px must be in plants; it only suggests
that the pressure is negative (below atmospheric). The CT theory
predicts a hydrostatic-pressure gradient, dPs/dx
0.01
MPa/m change in height, when there is no transpiration, i.e.:
|
(1)
|
where
is the density of water, g is the
acceleration due to gravity, and dh/dx is the
change in height per unit change in distance along a stem. The CT
theory must be augmented by the Ohm's law analog of water flow in
plants (van den Honert, 1948
) to predict dP/dx
values in stem segments with water flow rate > 0. The hydrostatic
pressure gradient in a stem segment will be augmented by a hydrodynamic
pressure gradient (dPh/dx)
when the water flow rate (f, in kilograms per second) in a
stem is >0, i.e.:
|
(2)
|
where R is the hydraulic resistivity of the stem
segment in megapascals per second per meter per kilogram.
Long-term measurements of Px in lianas
using the pressure probe have also been cited as evidence against the
CT theory. Benkert et al. (1995)
and Thürmer (1999)
report
pressure gradients of about
0.01 MPa m
1 at
night when f
0, but the CT theory appeared to be
challenged because dP/dx seemed to decline as
f increased. But this report must be discounted because the
authors committed two logical errors in their analysis of the data: (a)
a simple sign convention error and (b) the failure to integrate
Equation 2 over x to predict how
Px should change between two measuring
points at different heights.
The sign convention error occurred because their liana stem segments
were upside down, i.e. height decreased from physiological base to the
apex in the vines. The sign convention comes into Equation 2 when we
define the direction of increasing x to be from base to apex
and positive f to be for flow from base to apex. With this
sign convention it is obvious that dh/dx must be
negative. So the pressure gradient should start out as a positive
quantity at f = 0 then decline to 0 and then advance to
negative values as f increases. This is exactly what is
found in Figure 5 (Benkert et al., 1995
), so their experiments actually
provide strong qualitative support to the CT theory. However, their
experiments fail to provide strong quantitative support for the CT
theory because they did not measure the hydraulic architecture of the
lianas. Hydraulic architecture measurements would provide information
on hydraulic resistances within stems, petioles, and leaves. Such
information is needed to provide quantitative predictions of how much
Px should change between any two given
points in a shoot and how Px
differences should change with the transpiration rate. Even in
vertically oriented shoots, Px can
increase with height in some transpiring plants (Tyree, 1988
, 1997
) if
R decreases rapidly from large to small diameter branches.
Although the pressure bomb must be used with due consideration of what
it can and cannot measure, more recent evidence obtained with other
indirect methods supported the results obtained with the
Scholander-Hammel bomb and the CT theory (Holbrook et al., 1995
;
Pockman et al., 1995
). The discussion is still ongoing and alternative
mechanisms have been proposed, some of which are somewhat exotic
(Canny, 1995
) and argued to be erroneous (Tyree et al., 1999
).
The CT theory has been criticized occasionally over the past 100 years,
and there has been a search for alternative mechanisms. Criticism arose
from the striking fact that water under tension is in a metastable
state and should cavitate immediately when gas seeds are around.
Therefore, the xylem would be a quite vulnerable pipe (Milburn, 1979
).
The recent criticism does not completely exclude the CT mechanism.
However, it does claim that there have to be other mechanisms besides
CT in tall trees. Px values of only
0.5 MPa would be sufficient to lift a static water column only to a
height of 50 m, whereas the tallest trees can exceed 100 m.
Px would have to be more negative than
0.5 MPa in many common situations, e.g. when plants are in dry soils
or when dP/dx values must be very negative
because of large hydrodynamic gradients caused by high f or
R values (Eq. 2). Although the pressure probe has been used
to measure Px values down to
0.7 MPa
in extreme cases, little consideration has been given to theoretical
limitations of the pressure probe to measure negative pressures. How
vulnerable is the pressure probe to cavitation within the instrument?
How can the insertion of the pressure probe into vessels induce
cavitation in the vessel being measured?
From the above considerations it is clear that future tests of the CT
theory need to take into account quantitative aspects of the hydraulic
architecture of the plants being studied. In the present paper, we have
used 1- to 1.5-m-tall maize plants to test the validity of the CT
theory and the reliability of the pressure-bomb technique. We have also
examined the limitations of cell (xylem) pressure probes to measure
negative pressure.
 |
MATERIALS AND METHODS |
Plants
Maize plants (Zea mays L. cv Helix) were grown from
seeds in soil (sand:loam:peat, 1:2:1, v/v) in plastic pots (1.9 L; diameter: 150 mm; depth: 105 mm) in the greenhouse of the University
of Bayreuth (Germany). Plants were watered daily, and once a week were
supplied with nutrient solution containing 150 mM
K2HPO4, 150 mM
Ca(NO3)2, 200 mM
Mg(NO3)2, 100 mM
NH4NO3, 150 mM
(NH4)2SO4, and micronutrients. Experiments were conducted on 4- to 5-week-old plants that were 1 to 1.5 m tall. Plants were replaced after each experiment involving destructive sampling of a leaf tip.
Experimental Setup
A maize plant was brought from the greenhouse and set up for an
experiment as shown in Figure 1. The pot
containing the root was sealed in a metal pressure chamber, or
"root-bomb" (i.d. 185 mm, depth 305 mm), using rubber seals. Plants
were usually watered before placing them in the root-bomb, but in cases
where more negative Px values were
desired, the plants were not irrigated for 1 or more days prior to the
start of the experiment. Px was adjusted by changing the air pressure in the root-bomb. A pressure transducer (resolution ± 0.001 MPa) was mounted in the root-bomb to measure gas pressure.

View larger version (38K):
[in this window]
[in a new window]
|
Figure 1.
Experimental setup for measuring
Px using a cell pressure probe
(schematical). Px could be changed by
Pg to the root ("root-bomb") or by
changing light intensity, which affected transpiration. A cell pressure
probe was used to directly measure Px. When
a stable xylem pressure was observed, Pg was
increased in steps and then decreased again. Leaf tips were harvested
to measure water potential (balance pressure,
Pb) using a conventional pressure chamber
(shown in right corner). In some cases, leaf tips were covered with
aluminum foil (not shown), in other cases the leaf tip was allowed to
transpire. For further explanation, see text.
|
|
The direct measurement of xylem pressure using a cell pressure probe
was very sensitive to vibrations, which tended to cause cavitation and
failure of the experiment (presumably by air seeding where the tip
punctured the vessel). The cell pressure probe was mounted on a
manipulator (Leitz, Midland, Ontario, Canada) that was screwed on a
thick iron plate and placed on a heavy stone table. The iron plate was
also used as a magnetic stand for fixing a metal frame used to secure
the leaf while inserting the microcapillary into the vessel.
Measurement of Xylem Pressure
Xylem pressure was measured with an oil-filled cell pressure probe
rather than the water-filled xylem pressure probe used in many earlier
experiments (see "Results" for justification). The function of cell
pressure probes has been described in many earlier papers (e.g.
Steudle, 1993
; Henzler and Steudle, 1995
). The resolution of the
pressure transducer in the probe chamber was ±0.001 MPa.
Microcapillaries were made by pulling borosilicate glass capillaries
(i.d. approximately 0.5 mm; o.d. = 1 mm) on a microcapillary puller.
Tips were polished with a grinding machine (Bachhofer, Reutlingen,
Germany). Tip diameters (o.d. approximately 5 µm) and tip sharpness
(approximately 45° angle) were achieved during grinding.
Smaller-diameter tips cause less damage and are less likely to cause
cavitation during insertion into vessels, but hydraulic conductivity of
the tips declines with the fourth power of the tip diameter.
The half-time for pressure relaxation (pressure stability) of the
pressure probes was 3 to 10 s when the 5 µm were not plugged with cellular debris. The half-time for pressure relaxation measures the time for a pressure change beyond the tip of the pressure probe to
be registered in the body of the probe and requires the flow of
sufficient water across the tip to swell the volume of fluid in the
body of the probe. Microcapillaries were filled with low-viscosity
silicone oil (type CQ 240 D, Kulite, Leonia, NJ), but the tip was
filled with 3 to 4 µL of degassed water. During probing, no silicone
oil entered the xylem vessel probed. This was known because the
meniscus between silicone oil and water remained in the microcapillary.
The insertion of the probe was followed with a stereomicroscope. A
computer and a chart recorder were simultaneously used to record both
probe and root-bomb pressure.
The microcapillary was inserted about 0.2 m behind the leaf tip.
In some cases, the leaf tip was covered with aluminum foil to reduce
transpiration and to promote equilibration of water potential between
leaf tissue and xylem at the site of probe insertion. In other cases,
the leaf tip was allowed to transpire. Xylem vessels probed usually
were in ribs 2 or 3 counted from the midrib. The insertion of the
microcapillary into vessels was performed manually at an angle of 75°
to 90° between leaf blade and microcapillary. When the tip of the
microcapillary touched the rigid xylem wall, the tip bent,
"struggling" against the wall. The position of the microcapillary
was then adjusted (manipulator) to make it straight before puncturing
the vessel. The probe pressure was usually kept at an overpressure
during insertion in a vessel (0.02-0.05 MPa above atmospheric) by
slowly advancing the metal rod of the pressure probe. This helped
to prevent the formation of air bubbles in the tip and ensured that the
tip was not blocked during puncturing. When a negative pressure was
read with the probe, it was only possible to push the metal rod into
the probe and increase pressure. It was not possible to pull the rod
and decrease pressure without causing cavitation.
Criteria for Proper Measurement of Xylem Pressure
When the tip was blocked, a positive pressure pulse applied to the
probe (by means of the metal rod mounted into the probe chamber) did
not relax. When the tip broke during insertion, the probe pressure
rapidly returned to atmospheric pressure. The probe pressure returned
to subatmospheric (approximately
0.1 MPa) in less then 0.1 s
following cavitation, and then to atmospheric pressure in 10 to
100 s. Successful insertion of the tip in a vessel depended on
keeping it straight while pushing it forward. Several criteria for
successful insertion of the microcapillary into a vessel were used: (a)
the tip experienced bending against the vessel wall that was followed
immediately by a drop of probe pressure below atmospheric; (b) the
xylem pressure rapidly responded to changes in air pressure applied to
the root and to changes in light intensity; (c) water from the tip,
labeled with a dye, entered a single vessel following insertion.
Vulnerability to Cavitation (Tensile Strength) of Pressure
Probes
Direct measurement of negative pressure with probes requires that
there are no cavitation problems caused by the cell pressure probe
itself. In other words, when pressures down to
10 MPa are going to be
measured (Kolb and Davis, 1994
; Steudle, 1995
; Tyree, 1997
), probes
should not cavitate at pressures >
10 MPa. Usually, liquids withstand
high tensile stresses of up to several-hundred megapascals (Oertli,
1971
; Pickard, 1981
). However, in the presence of hydrophobic
surfaces, impurities, or fissures in the walls, cavitations may occur
sooner (Fisher, 1948
; Zimmermann, 1983
).
Seven different pressure probes were evaluated for vulnerability to
cavitation: five were fabricated in Würzburg (three water-filled and two oil-filled) and two were oil-filled probes fabricated in
Bayreuth. The pressure of the probe could be raised or lowered by
raising or lowering, respectively, the temperature of a sealed microcapillary. Sealed microcapillaries, about 80 mm long, were immersed in a water bath and the pressure changed about 0.1 MPa/1°C change in bath temperature for oil-filled capillaries. Water-filled capillaries changed pressure less with temperature because of the lower
thermal expansion of water versus silicone oil. Slow changes in
pressure (<2 kPa s
1) were achieved by
adjusting the temperature of the bath at the maximum cooling rate of
the refrigeration system while the microcapillary was totally immersed.
Rapid changes in pressure (>50 kPa s
1) were
achieved by lowering the microcapillary into a bath at 0.5°C over a
period of 10 to 30 s. In some cases the microcapillaries were
drawn on the puller and the tips ground, filled with water/oil, and
then sealed with glue. In most cases blunt capillaries were just
flame-sealed and filled with water and/or oil. The oil-filled capillaries still had water in the tip (10 mm); they were first completely filled with water using a 0.4-mm-o.d. syringe needle, then
back-filled with oil while holding the needle 10 mm back from the tip.
All solutions were partly degassed by vacuum before use.
Pressures at which a fracture of the liquid phase occurred were
recorded. Cavitations were evidenced by a rapid increase of pressure to
about
0.1 MPa (depending on the vapor pressure of the fluid),
followed by a gradual increase of pressure to 0 MPa as air came out of
solution to fill the cavity. In many cases the location of the
resulting embolism was observed at ×10 to ×30 magnification using the
stereoscope. Embolisms were removed by raising the temperature of the
fluid in the microcapillary until the pressures were above atmospheric,
which forced the gases back into solution. Probes could be put
repeatedly through cycles of positive pressure to remove embolisms and
negative pressure to induce cavitation.
Responses of Xylem Pressure to Root-Bomb Pressure and to Light
Most of the probing of leaf xylem was performed when root-bomb
pressure was atmospheric. When a stable xylem pressure was attained,
root-bomb pressure was increased in steps of 0.05 to 0.075 MPa and
kept constant at each level until a new constant xylem pressure was
established. Usually, steady pressures were attained after 8 to 12 min following a step change. The whole range of root-bomb
pressure included sufficient pressure for guttation to occur. The
root-bomb pressure was then decreased in steps to atmospheric.
Without root-bomb pressure, leaves were probed under a photosynthetic
flux density (PFD) of 150 µmol m
2
s
1 (measured at the leaf area probed with the
same orientation as the leaf blade). When a stable
Px was observed, light intensity was
increased by adding a second light, making the light intensity 200 µmol m
2 s
1. Data were
collected until a steady Px, and then
a third light was added (260 µmol m
2
s
1). The light sources used were two 400-W
lamps (Multi-Hi-Ace, Iwasaki, Japan) and one 400-W floodlight
lamp (Siemens AG, Frankfurt, Germany). Except for the study of
how light intensity affected xylem pressure and transpiration rate, the
other experiments were performed at 200 µmol
m
2 s
1 PFD. RH in the
lab was 60% to 75% and the air temperature was 20°C to 24°C.
Measurement of Balancing Pressure
(Pb) and Comparison with Xylem Pressure
Px
When a stable xylem pressure was observed with the pressure probe,
the leaf was harvested above the point of the insertion of the probe
and immediately wrapped in thin plastic wrap. An artificial petiole was
made by excising leaf-blade tissue from the midrib, while the leaf was
still wrapped in plastic. Then the balance pressure was measured in the
usual way with a pressure bomb. Gas pressure
(Pg) was increased in steps at a rate
<0.075 MPa min
1 between steps of 0.1 MPa or
less. When water appeared at the cut end of the midrib, the balance
pressure could be confirmed by lowering the bomb pressure by a <0.04
MPa until the water was sucked back into the xylem and then the balance
point was reconfirmed. Balance points were usually repeatable to
±0.005 MPa. The resolution of the pressure transducer used in the
pressure chamber was ±0.001 MPa.
Estimation of Transpiration under Different Conditions
In some experiments, the transpiration of plants was measured by
weighing the plant, including the pot and root pressure bomb. Transpiration was measured as a function of light (PFD) and the pressure applied to the roots. The weight of the plants and the root-bomb was measured on a balance scale with a resolution of ±0.01 g
at a maximum weight of 60 kg. The transpiration rate (W, kilograms per second) was computed from the weight change in 180-s intervals. During weighing, the tubing of the pressure bomb was arranged in a way so that it did not affect the measurement of small weight changes (as confirmed by adding small known weights to the
system). Pressure steps of 0.075 MPa were applied to the roots within
50 s and kept for about 10 min until the transpiration rate became
stable. The highest applied Pg was 0.6 MPa, which was then was decreased in steps. Similarly, light intensity
was changed (150, 200, and 260 µmol m
2
s
1) and the corresponding changes in the rate
of transpiration were measured. Transpiration became stable within 15 min after changing light intensity.
Hydraulic Architecture Measurements
The model used is shown in Figure 2.
The root was described by a single hydraulic resistance,
Rroot. The stem was divided into 10 segments divided at each node (segment resistance,
Rs). Stem resistances could be
measured only between midpoints of the internodes
(Rsi) because water-tight seals could
not be established at the nodes, hence the values used in the model for
any given stem segment was computed from half the basal internodal
resistance plus half the apical internodal resistance. The hydraulic
connection between the stem and base of the leaf sheath was described
by a leaf insertion resistance, Ri.
The leaf sheath and leaf blade were divided into segments 10 mm long
and each segment had two resistances: a xylem resistance,
Rx, and a mesophyll resistance, Rm. The xylem resistance is the
resistance of all the leaf vessels in parallel in the segment, and the
mesophyll resistance is the radial resistance for water flow from the
xylem vessels to the evaporative surface in the mesophyll air spaces of
the leaf. Component resistances in the model have been measured as
follows.

View larger version (32K):
[in this window]
[in a new window]
|
Figure 2.
Electrical analog used for the hydraulic
architecture of maize. The root is represented by a single resistance
Rroot and the stem is divided into 10 segments, the divisions being at the middle of the nodes, each with a
resistance Rs,1 to
Rs,10. A leaf insertion resistance is
assumed, RI,1 to
RI,10, is assigned to the resistance of the
complex vascular structure of the node. Each of the 10 leaves are
divided into 10-mm segments (up to 130 segments for the longest
leaves). Each leaf segment (for both leaf sheath and leaf blade) has an
axial resistance, Rx,i, for the resistance
of all the vessels in parallel and Rm,i, the
resistance of water movement from the vessels to the internal mesophyll
air spaces where water evaporates. The rate of evaporation from each
leaf segment is represented by a constant current source (circle with
arrow). This is justified because the normal range of leaf water
potential has little impact on the magnitude of the driving force for
water vapor diffusion ( X). The rate of evaporation is
given by Ai gL
X, where Ai is the surface
area of the ith leaf segment and
gL is the vapor diffusion conductance
(stomates, cuticle plus boundary layer). The guttation pathway is
represented by a diode. The diode permits liquid water flow (advance of
the meniscus) through leaf air spaces when the fluid pressure is >0.
The diode prohibits the movement of the meniscus (liquid flow) into the
cell wall (because of surface tension) when the fluid pressure in the
wall is <0 but still in a physiological range. These two electrical
components are best interpreted as a visualization of the boundary
conditions at the evaporative surface. Finally, a variable battery
represents the water potential of the soil plus the gas pressure
applied in the root pressure bomb ( soil + Pg).
|
|
Root Resistance
Transient measurements of root conductance,
Kroot = 1/Rroot, were made with a
high-pressure flowmeter (HPFM) (Dynamax, Houston). The theory of
operation of the HPFM when attached to roots is discussed in more
detail in Tyree et al. (1994
, 1995
). Tyree et al. (1995)
also showed a
linear relationship between F and
Pi and good agreement with the
Kroot measured by the pressure chamber method. The HPFM measures Kroot by
pushing water from the base of an excised root to the tips (opposite to
the normal direction of flow during transpiration). The shoot was
excised from the root about 0.05 m above the soil at the first
internode, hence the Rroot values
included the resistance of the first node plus half of the first
internode. The HPFM was attached to internodes (5-8 mm in diameter)
with the root system still in the pot. A water-tight seal between the
internode and the HPFM was achieved using the compression fittings
manufactured by Dynamax. Water pressure at the base was rapidly
increased from 0 to 0.5 MPa at a constant rate of 3 to 7 kPa
s
1 while measuring flow, F, and
applied pressure, Pi, every few seconds. The slope of the relationship between F and
Pi was taken as
Kroot, and
Rroot was calculated form the inverse
of Kroot.
Stem Resistances
The HPFM was first developed for measurement of shoot and leaf
resistances (for details, see Tyree et al., 1993
; Yang and Tyree,
1994
). Stem and leaf resistances were measured in the
quasi-steady-state mode with a constant applied pressure (P)
of 0.3 to 0.4 MPa. Stem segments cut at the midpoint of nodes could not
be sealed in the compression fittings without leaks, so resistances
were measured from stem segments cut at the midpoint of internodes with
one node in the center of the segment. The leaf sheath was removed from
the node. Constant pressure, P, was applied until a stable flow rate, F, was recorded and the resistance calculated
from R = P/F. During these
measurements, water flow followed two pathways: from the base of the
segment to the apex and from the base to the point of leaf insertion.
The flow through leaf insertion was measured by sealing the apical end
of the stem segment with cyanoacrylic glue, and the flow rate through
the leaf insertion, F', was measured at a constant pressure,
P'. The resistance of the leaf insertion was computed from
Ri = P'/F'. It
is not clear whether Ri should be
viewed as being in parallel with R or as a separate
resistance in series with R but emerging from the middle or
R. Values of Ri, measured when the apical internode was sealed with glue tended to be 11 to 14 times R (the resistance with both pathways open). The stem resistance, Rsi, was assumed to equal
to R
RI, which
probably underestimated Rsi slightly.
On the other hand, Ri included a small
portion of the stem resistance below the node and was probably overestimated.
At the other extreme, we could view
Rsi and
Ri as independent parallel resistors,
in which case Rsi should be equated to
R Ri/(R + Ri). Hydraulic models were computed
with both views of the resistance pattern with little difference in the
results, so we choose to use the former case to compute
Rsi and
Ri. The resistance values used in the
model equaled the resistance from the midpoint of the one node to the
midpoint of the adjacent node, and were computed from
Rs = (Rbsi + Rasi)/2, where
Rbsi and
Rasi are the stem resistance of
adjacent basal and apical stem segments, respectively. These
measurements were repeated for seven to eight segments from the base to
the apex of the plant. The two nodes at the top of the plant could not
be measured since the internodes were too soft to seal in the
compression fittings without crushing them.
Leaf Resistances
Leaf resistance could only be measured while attached to a node
because the HPFM could not be sealed to isolated leaf blades or leaf
sheaths. Most nodes were surrounded by the sheath of the leaf
immediately below. To isolate a single leaf we had to remove all leaves
to the apex without damaging the sheath so this could be done only on
leaves near the middle of the maize plants where the nodes were above
the sheath of the isolated leaf. The internode above the isolated leaf
was sealed with cyanoacrylic glue and the HPFM was connected to the
internode below the node to which the leaf was attached. The leaves
were between 0.9 and 1.2 m long (length of the sheath plus the
blade). The resistance of the entire leaf was computed from the applied
pressure, P, divided by the quasi-steady-state flow. Then
0.10 to 0.15 m of the leaf apex was excised. This increased the
flow and hence decreased the resistance of the remaining leaf. The
resistance was recorded and then the process of removal of 0.10 to
0.15 m of leaf apex followed by measurement of the residual
resistance was repeated until only the sheath remained. Then the sheath
was removed and the resistance of the remaining internode plus node was
recorded. The resistance was plotted against the length of leaf
remaining, and a computer program was written to fit the curve. Curve
fitting using the leaky cable model provided estimates of
Rm and
Rx per meter length of leaf.
Curve filling in the leaky cable model involved trial and error
selection of values of Rm and
Rx until a single pair of values provided estimates of leaf resistance that fit the entire curve of
resistance versus length of leaf remaining. The basic method of
solution is described in the discussion. A non-steady-state simulation
program was written to solve for the pressure and flow at each
resistance element. The simulation was iterated computationally until
the flow (f) into the base of the leaf equaled the sum of the flows out with a constant applied pressure (P) at the
base at which point the condition of steady-state flow has been met. The leaf resistance was then computed as P/f and
compared with experimental values. We found that a single pair of
Rm and
Rx values could predict the changes in
leaf resistance (P/f) as the leaf was
trimmed back from the apex. For a discussion of how the measurement of
Rm might differ from the effective
Rm during normal transpiration, see
Yang and Tyree (1994)
.
During HPFM measurements, the leaf air spaces filled with water and
water emerged both from stomates and through hydaothodes at the
leaf margins. A porometer (model 1600, Li-Cor, Lincoln, NE) was used to
estimate the rate of evaporation from the upper and lower leaf surfaces
during measurement of leaf resistance in the HPFM. This provided
information on the percentage of guttation that occurred through stomates.
 |
RESULTS |
Tensile Strength and Response Time of Pressure Probe
All pressure probes were cavitated 25 to 50 times. Usually, the
first few cavitations occurred at less negative pressures than
subsequent cavitations, but cavitation thresholds became repeatable
when probes were repeatedly cavitated with little time between. Typical
test runs for measuring the tensile strength are shown in Figure
3. Water-filled probes tended to cavitate at
0.6 to
0.7 MPa and oil-filled probes at
1.3 to
1.4 MPa. Embolisms were found to be near the metal rod (90% of the cases), suggesting that the most vulnerable air seeds were at the metal surface
of both water- and oil-filled probes. In 10% of the cases embolisms
were found at a plastic surface. Embolisms were never observed inside
the glass microcapillary and never at the oil/water interface (in the
microcapillary) of the oil-filled probes. In one instance, a sealed,
water-filled probe was taken down to
1.05 MPa (Thürmer et al.,
1999
), and in a few instances an oil-filled probe was taken to
1.6
MPa (B. Stumpf, personal communication), but these observations
are not representative of a "typical" pressure probe. Stable
negative pressures could not be held for more than a few minutes when
the pressure was within 90% of the cavitation threshold, but negative
pressures could be sustained for >1 h within 70% of the cavitation
threshold, i.e. about
0.5 and
1.0 MPa in water-filled and
oil-filled probes, respectively.

View larger version (23K):
[in this window]
[in a new window]
|
Figure 3.
Measurement of the vulnerability to cavitation of
cell pressure probes. Seven different pressure probes were cavitated 25 to 50 times each and these data are typical of many experiments
performed. Pressure was a function of the temperature of the sealed
microcapillary. In the case of oil-filled probes there was a linear
relation between pressure and temperature whereas for water-filled
probes the relations were non-linear (data not shown). A, Water-filled
pressure probes cavitated at less negative pressure than oil-filled
probes. Attempts to hold the pressure very near the threshold of
cavitation generally failed within seconds (first two attempts in
water-filled probe) and last attempt in oil-filled probe. B, Long-term
stability of an oil-filled probe that could hold 1.0 MPa for more
than 1.5 h without cavitation; water-filled probes could hold
0.5 MPa for the same period of time (data not shown).
|
|
Contrary to earlier suggestions, these data indicate that oil-filled
probes are more stable than water-filled probes. This means that
measurements with the cell pressure probe should be safe to
1 MPa.
This was the range measured in this paper. However, it should be noted
that the situation during an experiment (when the tip of the probe is
located in a vessel) was somewhat different from that during tests,
when the tip is closed, tending to reduce the useful range (see
"Discussion"). The tests also showed that cavitation in probes
could be reversed by the application of some overpressure for a short
period of time (Fig. 3). In the xylem, this is usually thought to be
the major mechanism by which repair of embolism takes place (by root
pressure or stem pressure in spring).
Figure 4 demonstrates how
Px measured with the pressure probe
responds to insertion into a vessel, to a rapid change in air pressure
in the root-bomb, and to a pressure pulse induced by rapid movement of
the metal rod into the pressure probe. When the probe was introduced
into a vessel, xylem pressure attained a stationary value after about
30 s (Fig. 4A). A pressure change in the roots was registered
rapidly in the Px of leaf vessels with
a response time of 5 to 20 s (Fig. 4B). Response times are related
to hydraulic resistances and capacities in the system, including the
pressure probe. Reduction of tip diameter increased its hydraulic
resistance and the half-time. By measuring the relation between rod
position and pressure when the microcapillary is sealed, we found that
about 100 nL of water must move into or out of the pressure probe per
megapascal change in pressure. (The 100-nL volume displacement is due
to the elasticity of the rubber seals in the pressure probe and not due
to the compressibility of water.) This volume displacement corresponds
to a water column 80 mm long in a vessel 30 µm wide, so a
considerable volume of water must flow through the leaf and into the
probe before a pressure change in the roots is registered in the probe.
In Figure 4C, a positive pressure pulse was produced by rapidly moving
the metal rod into the cell pressure probe. It can be seen that in this
case, the half-time was about the same as in Figure 4B.

View larger version (28K):
[in this window]
[in a new window]
|
Figure 4.
Test for proper location and functioning of cell
pressure probes in the leaf xylem of intact maize plants. In A, the
insertion of the cell pressure probe is shown followed by an adjustment
to a steady value of Px. In B, a pulse of
pneumatic pressure was applied to the root which was rapidly reflected
into a change of xylem pressure (half time: 8 s; see also Fig. 4).
In C, a typical pressure relaxation is shown following a pressure pulse
produced by the probe. Pressure relaxations exhibited short half-times
as well (half time of approximately 10 s in the experiment given
in the figure). Test B) indicates that the probe was able to rapidly
detect changes in xylem pressure. Test C indicates that the tip of the
probe was open and that its response time (half-time) for measuring
changes was short. It should be noted that the half-time measured in C
would represent an upper limit for the resolution in time of changes of
Px that could be detected.
|
|
Xylem Pressure versus Root-Bomb Pressure
A typical response of Px
(measured with the cell pressure probe) to
Pg is shown in Figure
5A. Px
responded differently to Pg depending
on the sign of Px.
Px changed almost as much as
Pg when
Px was < 0 and much less when
Px > 0. This was more clearly visualized when stationary values of
Px were plotted versus
Pg (Fig. 5B). Responses were linear in
both ranges of pressure. This experiment was repeated on 14 different
plants. When Px was < 0 (below
atmospheric pressure), the mean slope of the response was 0.846 (n = 14 plants; SD ± 0.086).
Although this slope was close to unity, it was significantly different
from unity and may be explained by a slight increase in the
transpiration rate of 15% as Px
increased from the initial value to 0 (see below). When
Px > 0 (above atmospheric pressure),
the slope was only 0.113 (SD ± 0.040). The air
pressure (Pg) required for guttation
to occur depended on factors such as soil water potential, hydraulic resistance of the plant, temperature, and relative humidity. Guttation started to occur at a certain threshold pressure
Px that decreased the hydraulic
resistance of the guttation pathway (hydaothodes and stomates).
The threshold pressure at which the slope changed was equal to or
somewhat higher than atmospheric pressure.

View larger version (20K):
[in this window]
[in a new window]
|
Figure 5.
Effect of Pg applied to
the root on xylem pressure, Px. A, At 0 (atmospheric) pressure, xylem pressure was
Px = 0.28 MPa. As the bomb pressure
was raised in steps, xylem pressure increased in steps, too. Responses
to bomb pressure were on the order of seconds (see Fig. 3). In the
range of xylem pressures of below atmospheric, there was a substantial
response of xylem pressure. This was strongly reduced when
Px was larger than atmospheric pressure and
guttation occurred. B, Effect of Pg applied
to a maize root on steady Px: plot of data
from an experiment such as that shown in A. White and black circles
represent the step-up and step-down of Pg,
respectively. Responses were linear in both the ranges of low and high
pressures. Guttation occurred when Px
attained a certain threshold (Px atmospheric pressure which was the reference). There was a nearly 1:1
response of Px:Pg
at low pressures ( Px slightly smaller
than Pg; slope = 0.852;
r2 = 0.998). At pressures where
guttation occurred, the slope was only 0.112. This indicated a
substantial reduction of the hydraulic resistance across the plant when
hydaothodes allowed the passage of water.
|
|
Leaf blades became translucent during guttation. When leaves were
examined under the microscope at ×200 with surface illumination, water
could be seen emerging from the stomates. When water emerged it tended
to spread out and form small pools on the leaf surface rather than
droplets. This guttation phenomenon could be reproduced using the HPFM.
Single leaves were perfused with the HPFM while simultaneously
measuring flow rate into the leaf, and the porometer was used to
measure the rate of evaporation of water from the surface pools on the
upper and lower surface of the leaves. Guttation through the stomates
accounted for 90% (SD ± 8%; n = 20) of
the flow into the base of the leaf; therefore, flow through hydaothodes was about 10% of the total flow into the leaves.
Effects of Light Intensity on Xylem Pressure
Light intensity may affect transpiration in two different ways.
First, at a given stomatal opening, an increase in light intensity increases leaf temperature and the water vapor pressure at the evaporative surface. This, in turn, increases the force driving for the
diffusion of water vapor across the stomatal pore. The other way in
which light intensity affect transpiration is that an increase of light
intensity will usually increase transpiration by increasing stomatal
width. Our results (Fig. 6) are in
agreement with the conventional idea of a continuous water pipe that
rapidly transmits changes in water potential across the plant (CT
theory). Light effects were reversible with similar time constants
(data not shown). Contrary to other findings (Benkert et al., 1991
), responses were substantial even at the relatively low light intensities used in the experiments presented in this paper (150-260 µmol m
2 s
1).

View larger version (24K):
[in this window]
[in a new window]
|
Figure 6.
Effect of light intensity on leaf xylem pressure.
In the typical experiment shown, light intensity was 150 µmol
m 2 s 1, when a xylem vessel of the leaf was
probed (arrow). It was then increased in steps to 200 and 260 µmol
m 2 s 1 (arrows). It can be seen that there
was a substantial and rapid response of Px
to light intensity that correlated with a higher transpiration rate as
measured by weighing the plant (data not shown). Note that the
half-time for attaining a steady Px after
puncturing was about 5 s and about 20 to 50 s following a
change in light intensity.
|
|
Comparison between Px and
Pb
A total of 65 leaves was used for the comparison of
Px with
Pb, with stable
Px ranging between
0.11 and
0.73
MPa. Comparison showed a good agreement between
Px and
Pb (Fig.
7). For transpiring leaves, the slope was
0.967 (r2 = 0.9964; n = 36 leaves); for non-transpiring leaves, the slope was 0.984 (r2 = 0.9988; n = 29 leaves). Both slopes were not significantly different from unity
(t test; P = 0.05). These statistics were based only on the Px values that were
stable during the measurements. Forty-two other experiments were
performed in which cavitations occurred before stable
Px values could be observed, including Px values down to
1 MPa. These
values were not included in Figure 7 because
Pb is an equilibrium measure, whereas
the pressure probe is capable of both dynamic and equilibrium measures
of Px.

View larger version (24K):
[in this window]
[in a new window]
|
Figure 7.
Comparison between Px
and Pb measured in the same leaf (Fig. 1).
Each symbol represents measurements of Px
and Pb for a different leaf (total: 65 leaves measurement on 65 plants). , Leaf tip was covered with
aluminum foil to prevent transpiration (slope = 0.984); ,
uncovered leaf that was allowed to transpire (slope = 0.967).
The results indicate that the pressure probe and the Scholander-Hammel
bomb measure similar values.
|
|
The dynamic situation is characterized by water flow through the leaf
and gradients of water potential and
Px. These gradients disappear when
leaves are harvested for Pb
measurement. A stable Px value
indicates that the leaf had approached steady-state evaporation with
pressure gradients that are time independent. We would expect the
offset between Px and
Pb to be proportional to the
steady-state transpiration rate. In our experiments transpiration was
approximately constant and Px was
changed by adjusting Pg; therefore, we
expect and in fact found a constant offset between
Px and
Pb. This is in contrast to
less-controlled experiments in which
Px varied because of changing
transpiration in leaves (Fig. 1; Melcher et al., 1998
). Melcher et al.
(1998)
found better agreement between Px and
Pb in non-transpiring leaves than when
the leaves were transpiring, which is fully consistent with the CT theory.
Effect of Root-Bomb Pressure on Transpiration Rate
Transpiration rate very much depends on environmental conditions
(air temperature, relative humidity, wind speed, etc.) and the plant
itself (water status, leaf temperature, stomatal opening, leaf area,
etc.). The transpiration rate should increase with increasing root-bomb
pressure, because increasing Pg
increases the water potential of the shoot and improves its water
status. This was observed in the experiments, and a representative plot is shown in Figure 8. Average evaporative
flux density (E) measured by weighing ranged from 0.5 to
6 × 10
6 kg s
1
m
2 in five replicate experiments. Higher
Pg resulted in higher transpiration rates. Effects were reversible, i.e. when
Pg was decreased, transpiration also
decreased. Pg was changed in steps of
±0.075 MPa within 50 s. After a step change in
Pg, 1 to 2 min were required to attain a new stable Px. Thus, a time period
of 9 min was sufficient to attain a stable transpiration rate.

View larger version (12K):
[in this window]
[in a new window]
|
Figure 8.
Effect of increased Pg
on E. , Increasing pressure; , decreasing
pressure. Each point is the mean of three 3-min readings and pressure
was changed every 10 min. The light intensity for these experiments
were less than for Figure 5 because, to get accurate weights, the plant
had to be moved so that it did not touch nearby objects; the lights
were not moved with the plant. Consequently the initial
E values were lower than for most other experiments.
|
|
When E was plotted against
Pg, the relationship was linear when
Pg > 0.2 MPa, which corresponded to
the pressure at which guttation began near the base of the maize shoot.
The slope with guttation (Pg from 0.2-0.6 MPa)
was 7.5 to 9.4 × 10
5 kg
s
1 m
2
MPa
1 in five replicate experiments. The slope
was significantly less, 5 to 8 × 10
6, for
Pg < 0.2. Our findings are in
agreement with other reports of an increase in E with
increasing root-bomb pressure (Saliendra et al., 1995
; Fuchs and
Livingston, 1996
).
Hydraulic Resistance
The hydraulic resistance of leaves was maximal in whole leaves and
decreased as progressively more tissue was removed from the apex. A
typical plot of leaf resistance versus the length of leaf remaining is
shown in Fig. 9. These data were fitted
to a leaky-cable model of water flow through leaves producing a good fit in all six leaves measured. The Rx
of a 1-m length of leaf blade was 6 × 104 ± 1 × 104 (SD) MPa
s
1 kg
1. The
Rm of a 1-m length of leaf was
4.5 × 104 ± 0.8 × 104 (SD) MPa
s
1 kg
1. To simulate the
change in resistance after removal of the leaf sheath we had to assume
that the xylem and radial resistances were about 1.8 times that of the
leaf blade. A summary of root and stem resistances is shown in Table
I. The leaf insertion resistances were
about 10 times the resistance between adjacent nodes.

View larger version (14K):
[in this window]
[in a new window]
|
Figure 9.
Theoretical ( ) and measured ( ) hydraulic
resistance of a single maize leaf as a function of the leaf length
(including the sheath). The leaf was progressively cut back from
the tip and the resistance was recorded after each cut. The leaf
parameters for the theoretical curve were:
Rx = 6.0 × 104 MPa
s 1 kg 1 m 1 and
Rm = 4.5 × 104 MPa
s 1 kg 1 m 1 and a stalk plus
node resistance of 1.68 × 104 MPa s 1
kg 1. The Rx and
Rm values of the leaf sheath had to be set
to 1.8 times the values for the leaf blade to fit the last two points.
|
|
View this table:
[in this window]
[in a new window]
|
Table I.
Summary of Rroot, Rsi, and
Ri
Maize plants had 10 nodes numbered from 1 (base) to 10 (apex).
Rroot includes node 1 because the HPFM could not be sealed
to the root below node 1. Resistances for nodes 9 and 10 could not be
measured because the tissue was too soft to mount on the HPFM
connector, but the values used for modeling purposes are given in
square brackets. All values are given as means ± SE;
n = 6.
|
|
 |
DISCUSSION |
Pressure probes as they stand have a fairly limited tensile
strength. They may cavitate at a pressure that is considerably less
negative than that proposed to exist in the xylem of some transpiring
plants (
10 MPa according to indirect measurements with the pressure
bomb; Kolb and Davis, 1994
). To our knowledge, there have been no
attempts to measure the vulnerability to cavitation in oil-filled
versus water-filled probes nor any suggestions on how to improve probes
based on these measurements reported in the literature. Oil-filled
probes seem to be superior to water-filled probes, which cavitate at
0.7 versus
1.4 MPa, respectively. Therefore, each type can be used
only in its respective range. Immediate improvements in technique would
result by switching to oil-filled probes. Another useful improvement
would be to replace the metal rod with a rod made out of a material
with better adhesive property with the surrounding oil. But as
materials are changed in probe design, both water- and oil-filled
versions should be tested. With some material, the higher surface
tension of water may prove advantageous because of the role of surface
tension in air seeding.
With the pressure probe technique as it stands it is not possible to
test predictions about the existence of very negative values of
Px as predicted from experiments with
the pressure bomb and other techniques. The minimum
Px measured in punctured vessels were
somewhat smaller than those found with sealed probes, probably because
of probability of air seeding through the seal in the annulus between
the outer surface of the micropipette and the wall of the punctured vessel.
From the minimum value of pressure, which could be obtained in sealed
pressure probes, we may estimate the maximum diameter (2r)
of the air (gas) seeds causing cavitation. This would be given by
P = 2
/r, where
P
represents the pressure difference of the fluid minus the vapor
pressure of the fluid and
is the surface tension of the fluid
(
= 0.025 Newtons m
1 in an
oil-filled probe and 0.073 in a water-filled probe at 20°C). In an
oil-filled probe the critical diameter is 76 nm when
P = 1.3 MPa and for a water-filled probe the
critical diameter is 420 nm when
P = 0.7 MPa, i.e.
diameters of gas seeds were less than or equal to the wavelength of
visible light (400-800 nm). Cavitations are not thought to arise in
the bulk liquid phase (water, oil). Cavitations are thought to arise at
air seeds harbored in solid/liquid interfaces. Surfaces have to be
clean enough to prevent gas-seeding (Fisher, 1948
; Briggs, 1950
;
Zimmermann, 1983
; Steudle, 1995
). Once the probe is inserted into a
vessel another locus of air seeding is the annulus of space between the
cell wall of the vessel and the outer surface of the probe.
A thin layer of water with a meniscus may exist at this annulus. The
thickness of this annulus,
, may prove to be the ultimate limitation
to the pressure probe techniques. Air seeding through the annulus would
occur whenever
P > 2
/
(assuming
is much less than the diameter of the hole in the vessel created by the probe).
Assuming this annulus harbors an air/water meniscus, the thickness of
the annulus must be less than 296 nm when
Px =
1 MPa, which is the most
negative pressure measured in this study. Therefore, future experiments
designed to measure pressure down to
10 MPa would succeed only if the
pressure probe is improved and the annulus is <30 nm thick, i.e. about
one-third the thickness of a cell membrane.
Contrary to earlier reports, we find many results immediately
consistent with the CT theory, e.g. the quick response of pressure measured in vessels of maize leaves to changes in root-bomb pressure (Fig. 4) and to changes in light (Fig. 6). We also found good agreement
between the pressure bomb and the pressure probe (Fig. 7). Our results
more than double the range of agreement between the pressure probe and
the pressure bomb, i.e. 0 to
0.7 MPa (Fig. 7) versus 0 to
0.3 MPa
in Melcher et al. (1998)
. Other results, such as Figure 5 and the
disagreement between Px and
Pb in transpiring leaves reported by
Melcher et al. (1998)
, requires a more detailed analysis of the
hydraulic architecture to explain.
A computer model was written to solve for the pressure drop across the
network of resistors shown in the hydraulic architecture model of a
maize plant (Fig. 2). The method of computation was identical to that
used previously (Tyree, 1988
) for the dynamic solution of water flow
through large trees. The capacitance associated with each resistance
element was not measured, so realistic non-steady-state solutions could
not be computed, but steady-state solutions are independent of
capacitance. Each resistance element (R) was assigned an
arbitrary capacitance value (C) such that RC = 1 s. The dynamic solutions were iterated with time steps of
dt = 0.5 s until the rate of water flow into the
root resistance equaled the sum of water flow out of all leaf segments,
which is the condition that defines steady state. The flows at the root
and leaf boundaries were determined by the boundary conditions.
The boundary condition used for the root was a pressure equal to the
soil water potential plus Pg. The
boundary condition at the surface of the leaf segments depended on the
pressure at the surface. When the surface pressure was negative, the
boundary condition was a constant evaporation rate:
Ai gL
X, where Ai is the
surface area of the ith leaf segment,
gL is the vapor diffusion conductance
(stomates, cuticle plus boundary layer), and
X is the driving force
on vapor diffusion. The value of Ai
gL
X was set at
E* + dE*/dPg
where dE*/dPg was the rate
of increase of evaporative flux density with
Pg observed in this study before guttation. Guttation starts when the pressure at the surface equals or
slightly exceeds 0 (atmospheric pressure). Therefore, the boundary condition at the evaporative surface was changed from E* + dE*/dPg to pressure = 0 when guttation occurred.
The program listing was originally written in Turbo Pascal and later
updated to a Windows 95/98 version of Pascal (Delphi 4.0). Space does
not permit printing the full program in this paper but a copy of the
program will be provided upon written request to the corresponding author.
The model successfully predicted the observed dependence of xylem
pressure at the pressure probe (Px) on
whole plant evaporative flux density (E) and the dependence
on changes in root-bomb pressure (Pg),
Figure 10, A and B. The model also made
realistic predictions of the gradients of
Px throughout the entire shoot. Figure
10C (left and right axis) shows the predicted gradients of
Px that occur at the PFD of 200 µmol
s
1 m
2 (our experimental
conditions) and at 1,200 µmol s
1
m
2 (typical greenhouse conditions on a sunny
day), respectively.

View larger version (33K):
[in this window]
[in a new window]
|
Figure 10.
Theoretical output of the hydraulic architecture
model. A, Computed Px in leaf 5 (0.25 m from
the tip) versus Pg (= root-bomb pressure);
compare to data in Figure 5B. B, Computed water flow rate put unit leaf
area of maize shoot versus Pg; compare to
data in Figure 8. C, Computed profile of Px
versus distance in maize stalk and leaves when
Pg = 0 and all leaves have an
evaporative flux density of E = 2.6 × 10 5 kg s 1 m 2 (= 1.44 mmol
s 1 m 2) on left axis and for
E = 13 × 10 5 (7.2 mmol
s 1 m 2) on right axis, which corresponds to
the probable value in a greenhouse. D, Computed profile of
Px versus distance in a maize stalk and
leaves when Pg = 0.6 and guttation
occurred from all leaves. Guttation rate at each point along the leaves
is in proportion to Px, so most guttation
occurred near the base of the leaves and mostly from the lowest
leaves.
|
|
The hydraulic architecture model was able to simulate the change of
slope as well as the approximate value of the slopes between positive
and negative values of Px (Fig. 10A).
The evaporative flux, E*(leaf area), and liquid flow rate
should be equal under steady-state conditions. The slope for
Px < 0 would have been 1.0 had the
steady-state E remained constant, because then the liquid
flow rate from the root to the evaporative surfaces would have been
constant and the difference in pressure would have therefore been
constant as well. Therefore, an increase in root pressure would
necessarily have caused an equal increase in
Px. Direct measurements of
E*(leaf area) demonstrated an increase in evaporative flux.
An increase in E*(leaf area) should cause a decrease in Px in the leaf, and an increase in
Pg should cause a 1:1 increase in
Px; when both effects are
superimposed, a slope < 1 between Px and
Pg would result. Model results confirmed that an
increase of E*(leaf area) of about 15% explained a slope of
0.85, while Pg changed from 0 to 0.3 MPa.
This 15% increase in evaporative flux was input into the model as the
boundary condition at the leaf surface, but doing so does not explain
the increase in evaporation rate. Since the evaporation rate = A gL
X, the increase
must be caused either by an increase in
gL or an increase in
X.
The increase in
X can be estimated from the environmental
conditions. Air temperature and humidity were about 22°C and 65%,
respectively, so the water vapor pressure in the air must have been
1.719 kPa (computed from vapor pressure tables). During the experiment
the leaf water potential changed from
0.3 to 0 MPa, which will affect
RH and hence vapor pressure at the evaporative surface (equation 2.21 of Nobel, 1991
). Given a leaf temperature of 22°C, the vapor
pressure at the evaporative surface must have been 2.639 and 2.645 kPa
at
0.3 and 0 MPa, respectively, and the increase in
X
would have been 100*(2.645
2.639)/(2.639
1.719) = 1.4%. Consequently, the increase in evaporative flux must have been
caused by an additional 13.6% increase in
gL and this would explain the slope of
0.85.
The dramatic fall in slope from 0.85 to 0.11 (see "Results") was
somewhat overestimated by the model, which gave a change in slope from
0.832 to 0.06 (Fig. 10A). This dramatic change in slope is caused in
part by the dramatic rate of increase in water flow across the root and
basal portion of the shoot due to guttation (Fig. 10B). The model
predicted that most of the guttation was confined to the base of the
shoot (Fig. 10D) and that the guttation started while
Px was still negative (where the
pressure probe was located) and Pg = 0.2 MPa (Fig. 10B and other data not shown). Therefore, the water flow
was "short-circuited" by the guttation pathway at the base of the
shoots (lower diodes in Fig. 2). The pressure probe measured only the
much reduced pressure required to drive the reduced water flow rate
from the probed vessel at Px > 0 across the mesophyll resistors (Rm)
adjacent to the probe in Figure 2. The model could be fine-tuned to
exactly simulate the observed change in slope by making the values of
Rx and
Rm larger in the lower leaves than in
the upper leaves, but we could not confirm this experimentally. The
morphology of the shoot permitted measurement of leaf resistance (Fig.
9) only in the largest leaves (leaf 5 or 6) near the middle of the shoot.
The steady-state gradients of Px are
shown in Figure 10C for low and high evaporation rates (read from left
and right axis, respectively). These curves provide useful insights
into the reason for the deviation between
Px and
Pb in transpiring leaves. The relationship (in this paper) between
Px and
Pb was very linear with little
dispersion of data points about the lines in both transpiring and
non-transpiring leaves; the offset between the two lines was also small
(about 0.01 MPa in Fig. 7). In contrast, the dispersion of data and the
offset was much more in similar work reported by Melcher et al. (1998)
.
The differences can be explained by the hydraulic architecture model
and by differences in experimental design. In our experiments the
transpiration rate was lower than in Melcher (1998)
; and in our
experiments the leaf tip (0.2 m) just above the probed vessel was
harvested for comparison of Px and
Pb; conversely, Melcher et al. (1998)
harvested adjacent leaves in the plant.
The model predicted rather minor gradients of
Px in the leaf tip when
E = 2.6 × 10
5 kg
s
1 m
2 (Fig. 10C, read
from left axis), and much bigger gradients (Fig. 10C, read from right
axis) when E was 5 times more. The model also predicted
large differences in Px between leaves
even when E was the same in every leaf; therefore, the
experiment of Melcher et al. (1998)
was not well designed to compare
Px and
Pb because the presumption that
adjacent leaves should have nearly identical pressures is clearly
wrong. The amount of variation in Px
between adjacent leaves is likely to be even more than shown in Figure 10C (right axis), because this model was computed assuming the same
E value for every leaf. In reality, E will be
approximately proportional to the adsorbed light energy and the amount
of light absorbed by each leaf will vary widely between leaves. Light
interception will be equal to the incident radiation times the sine of
the angle between the incident rays and the surface of the leaf plane. In addition to this, some leaves will be shaded by other objects (equipment, stalks, leaves, etc.). We recommend that all future comparisons of Px to
Pb be done on the same leaves and that
care be taken to document the light intensity on the measured leaf.
Recently, Thürmer et al. (1999)
report an extended study of
Px and simultaneous cell turgor pressure
(Pt) measurements in mesophyll cells of the
leaves of a liana, Tetrastigma voinierianum. Everything
presented in that paper is qualitatively consistent with the CT theory,
even though the authors failed to draw this conclusion. The authors
drew a rather startling and incorrect inference that the minimum
attainable Px value is determined by Pt, i.e. when
Pt reaches 0 there is no longer a
stable equilibrium between Pt and
Px. Thürmer et al. (1999)
write
"Considering that the xylem pressure is determined by the turgor
pressure (and vice versa), the xylem pressure of the liana could not
drop to... less than
0.4 MPa because this pressure corresponds to
zero turgor pressure." The water potential of a vessel should be
nearly equilibrated with the water potential of an adjacent living
cell. Changes in cell water potential are driven mostly by changes in
Pt with little change in osmotic
pressure (
) (see fig. 1 in Tyree, 1999
). Therefore, Pt and
Px, should increase and decrease
together in nearly a 1:1 relationship when the cell osmotic pressure
(
) is nearly constant; this has been confirmed quite conclusively in
Thürmer et al. (1999)
.
The minimum Px is limited only by
vulnerability to air seeding and not by the nature of the equilibrium
(or lack of equilibrium) between water potential in the xylem
(determined mostly by Px) and water
potential in living cells (determined by
Pt
). Clearly, the water
potential of living cells can drop below the turgor loss point (Tyree
and Jarvis, 1982
) and Px will be
approximately in equilibrium with cell water potential until the limit
of pressure by air seeding is reached. In woody living cells,
Pt could swing to negative values as
water loss progresses past the turgor loss point, because the lignified
wall prevents cell collapse. Even if these rigid living cells cavitate,
equilibrium will continue to exist between the cavitated living cells
and the xylem vessels. In soft mesophyll cells,
Pt will fall to and remain near zero with water loss beyond the turgor loss point. Mesophyll cells will
collapse and decrease in volume because the cellulose is not lignified,
and water potential will equal
beyond the turgor loss point, as is
easily confirmed by pressure-volume curves (Tyree and Jarvis, 1982
).
Cavitations generally begin in some vessels near the turgor loss point
in many species, but many vessels remain functional to much more
negative values of Px; however, there is no cause-and-effect relationship between loss of turgor in living
cells and the start of cavitations.
In conclusion, there is strong evidence in favor of the CT theory of
water movement in plants. The CT theory combined with the Ohm's law of
quantification of the transport process provides a very robust model
capable of explaining all of the observations in this paper. There is
also good agreement between the pressure bomb and the pressure probe
under well-defined conditions. Large gradients in
Px in transpiring leaves have been
known for a long time (Begg and Turner, 1970
; Turner and Long, 1980
;
Turner, 1981
). The existence of such gradients does not invalidate the
pressure bomb technique. The pressure bomb can measure only the
equilibrium Px of a leaf in a
non-transpiring state, which necessarily follows after a leaf is
enclosed in a dark and humid chamber. However, most people use the
pressure bomb to estimate the average leaf water potential of
transpiring leaves. The pressure bomb is ideally suited to do this
because, once a transpiring leaf is harvested and mounted in the
pressure bomb, the gradients dissipate in the leaf and equilibrium
Px that results is an "average" of
the gradients of water potential that existed in the leaf prior to
excision. The only exception to this generalization would be expected
in leaves with high solute concentrations in the xylem fluid. In general Px = "average" water
potential of the living cells plus the "average" osmotic pressure
in the xylem fluid. Some of the theory on how water potentials average
has already been worked out (Tyree and Hammel, 1972
; Tyree, 1981
). The
xylem pressure probe is ideally suited to measure
Px at any given point in a leaf and
thus could be used to directly measure the gradients within the leaf
and to quantify how the water potential gradients within the leaf
"average out" to achieve the balance pressure.
 |
ACKNOWLEDGMENTS |
M.T.T. thanks the Humboldt Foundation (Bonn) for a research
award (Humboldt-Forschungspreis) and Prof. U. Zimmermann for use of several water-filled and oil-filled pressure probes during tests of
cavitation thresholds. We acknowledge the expert technical assistance
of Burkhard Stumpf (Lehrstuhl für Pflanzenökologie, Universität Bayreuth).
Received April 27, 1999; accepted July 23, 1999.